TELECARE M

Nurse Call System ***OUT OF ORDER***

  • Pequeños módulos Centrales.
  • Arquitectura descentralizada.
  • Software de Gestión en entorno Windows con tecnología web.
  • Sistema de canales múltiples de voz.
  • Tecnología LON.
  • 2 variantes: con comunicación y sin comunicación.
  • Integrable con mensajería Ascom e IP-Dect.
  • Posibilidad de integración con otros sistemas.

  • 8 wires (2 wires for LON bus + 6 wires for communication bus).
  • Maximum of 64 equipment per bus.
  • 500 m max per bus.
  • Requires end of line resistance.
  • Increase the system by LBR repeaters.

Centralized installation
  • Small bus network.
  • A single Power Supply.
  • Longer length cables to peripherals.

Decentralized installation
  • IAM units next to peripherals.
  • Reduced cabling to peripherals.
  • Extensive network of bus.
  • Power Supply decentralized.

Individual distribution of peripheral
  • General call with a unique address.
  • Identification per bed.
  • Programming for multiple rooms.

Room Bus Connection
  • Multiple rooms can be connected to a bus.
  • Reduction of IAM units.
  • Extensive network of bus.
  • No identification bed.
 

Nurse Station Display
  • Displays all the system calls or group selected calls.
  • Information on the status of the instances of nurses entir system or selected groups.
  • Audible calls.
  • LCD 2 x 16 lines.
  • White or gray.
  • Dimensions 184 x 108 mm.

IP Corridor Display (Requires ISC2)
  • Versions of 6 or 12 characters.
  • Single or double side.
  • Text in 3 colours (red, green, amber).
  • For wall or ceiling mount.
  • Directly to bus LON telecare
  • Dimensions:
    • 6 characters, single side: 350 x 135 x 38 mm
    • 6 characters, double side: 350 x 135 x 76 mm
    • 12 characters, single side: 650 x 135 x 38 mm
    • 12 characters, double side: 650 x 135 x 76 mm
  • Power: 24 VDC
  • Max. Power Consumption:
    • 6 characters, single side: 14,4 W
    • 6 characters, double side: 24,0 W
    • 12 characters, single side: 21,8 W
    • 12 characters, double side: 38,4 W

IP Corridor Display (Requires ISC3)
  • Versions of 6 or 12 characters.
  • Single or double side.
  • Text in 3 colours (red, green, amber).
  • For wall or ceiling mount.
  • Directly to LAN
  • Dimensions:
    • 6 characters, single side: 355 x 135 x 38 mm
    • 6 characters, double side: 355 x 135 x 76 mm
    • 12 characters, single side: 655 x 135 x 38 mm
    • 12 characters, double side: 655 x 135 x 76 mm
  • Power:         24 VDC o PoE (ver consumos)
  • Max. Power Consumption:
    • 6 characters, single side: 7,9 W
    • 6 characters, double side: 15,0 W
    • 12 characters, single side: 11,3 W

Room Message Display
  • Display 6 digits + symbols.
  • Buzzer incorporated.
  • Navigation between messages.
  • Interactive Messaging.
  • You can add communication module response NSAM.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 148 x 17.
 

Speech Answer Module
  • Unit to answer calls made from rooms.
  • Microphone and 2 speakers for hands-free communication.
  • Displays can be connected to Room.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.

Communication Module
  • Microphone and 2 speakers for hands-free communication.
  • Interactive Messaging Compatible with Ascom.
  • can be connected with Room Displays or modules with / without socket.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.

Modules without connector
  • Can be 1, 2 or 3 buttons.
  • To call, assistance and cancel.
  • You can add communication module.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.

Modules with connector
  • Can be 1 or 3 buttons.
  • To call, assistance and cancel.
  • Optional light control circuit.
  • Socket Safe Release for mini handset.
  • You can add communication module.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.
 

Socket for Medical Rail
  • Recessed into Medial Rail.
  • Socket Safe Release for mini handset.
  • Optional light control circuit.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.

Mini Handset
  • With socket Safe Release.
  • Sterilized and water resistant.
  • Red button call.
  • Version with 2 yellow buttons to control lights.
  • Available with cable 2.5 m or 6 m.
  • White or gray.
  • Dimensions WxHxD (mm): 151 x 53 x 22.
 

 
Patient Handset with Light control and Sundblind control
  • With socket Safe Release.
  • Sterilized and water resistant.
  • Red button call.
  • 2 light switching buttons.
  • Sunblind control buttons *.
  • Integrated bright LED flash light
  • Service button.
  • White or gray.
  • Dimensions WxHxD (mm): 151 x 53 x 22.
* Only available for telecare IP, it requires module 660381.

 
Multimedia Patient Handset with 14 buttons
  • With socket Safe Release.
  • Sterilized and water resistant.
  • Red button call.
  • 2 light switching buttons.
  • Sunblind control buttons *.
  • Speaker and microphone for support speech in teleCARE IP and teleCARE M.
  • IIntegrater Infrared TV control and an integrated FM radio receiver
  • Radio/TV control buttons and jack socket for audio (sterea with telecare IP)
  • Integrated bright LED flash light
  • Service button.
  • White or gray.
  • Dimensions WxHxD (mm): 151 x 53 x 22.
* Only available for telecare IP, it requires module 660381.

Bathroom Module
  • With 1 button for call, or 3 buttons to call, assistance and cancel.
  • With 2 meter cord handle.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.
 

Mounting Plates
  • Needed to fix to wall the telecare units.
  • Three types: single, double and triple.
  • White or gray.
  • Dimensions WxHxD (mm):
         91 x 91 simple.
         91 x 184 double.
         90 x 270 triple.

Duty Selector
  • Select up to 10 pre-programmed tasks.
  • Includes buzzer.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.
 
 

Corridor Unit
  • For identification of the caller (bedroom or bathroom).
  • Has Led units sleep up to 3 colors (colors available: red, green, yellow)
  • White or gray.
  • Dimensions WxHxD (mm): 160 x 123 x 57.
 

Entretainment Module
  • Slider up to 8 channels of music.
  • Buttons to raise and lower the volume.
  • Integrated with patient call units.
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.

Acoustic Activation Call Module
  • Call activation for noise or acoustic signal.
  • LED indicator for detecting sound.
  • LED activation of the direction of the voice channel.
  • Highly sensitive microphone for sound detection and the two-way voice communication
  • Two high quality speakers for voice
  • Set variable level and duration of sound
  • White or gray.
  • Dimensions WxHxD (mm): 91 x 91 x 17.
 

Intelligent Address Module IAM
  • 10 adrress without communication for rooms.
  • Possibility to install SPB board to have the first 8 communication address.
  • Programmable as SBM (spoken communication manager) and SMM (supervisor modules).
  • Dimensions WxHxD (mm): 135 x 270 x 60.
  • Power supply: 24 VDC.

Telephone Line Module TLM
  • Interface for telephone communication, with voice communication of teleCARE system.
  • 2-channel communications per module.
  • Requires 2 analog telephone extensions 48 V.
  • Dimensions WxHxD (mm): 135 x 270 x 60.
  • Power supply: 24 VDC.

Repeater Module LBR
  • Expands the system in case of exceeding the bus over 500 meters or more than 64 system modules.
  • Expand the system up to 512 modules.
  • Buses galvanically isolates.
  • Dimensions WxHxD (mm): 135 x 270 x 60.
  • Power supply: 24 VDC.

Central module ISC3
  • Enables wide management of teleCARE System.
  • Necessary if use corridor displays, NSS, IAM SMM and integration with paging systems, IP-Dect and Wandering.
  • Requires licenses according needed.
  • Dimensions WxHxD (mm): 220 x 199 x 44.
  • Power supply: 230 VAC.

Nurse Station Server Module NSS
  • Retrieves call and system information from the ISC and publishes this information on its internal NSS website.
  • Allows access via web browser, the following information:
    - Scheduling, register assignment, Printing records and events,
    - Consultations Archives, Search Options, Filter.
  • Requires licenses according needed.
  • Enrackable optional 19" 1U.
  • Dimensions WxHxD (mm): 220 x 199 x 44.
  • Power supply: 230 VAC.

Serial LON Adapter Module SLA
  • Allows remote access, management and maintenance of teleCARE M.
  • Connects to the teleCARE LON bus, and a serial port modem or PC with external access control.
  • Dimensions WxHxD (mm): 94 x 157 x 36.
  • Power supply: 24 VDC.

Serial LON Adapter Module SLA, Second Version
  • Elise 2 module.
  • Allows remote access, management and maintenance of teleCARE M.
  • Connects to the teleCARE LON bus, and a serial port modem or PC with external access control.
  • USB port for PC connection.
  • Dimensions WxHxD (mm): 135 x 270 x 60.
  • Power supply: 24 VDC.

NSS

ASSIGNATIONS

CALLS NOTIFICATION

ACTIVE CALLS

LIST

EXPORT